Instrumentation Lab Manual

Sr.no

          practicals          

1

Introduction  of Colorimeter

 

2

To find out  of KMNO4  using colorimeter

3

To find out  of K2Cr2O7 using colorimeter.

 

4

To find out unknown concentration of sample using colorimeter.

 

5

Introduction of UV visible Spectrophotometer

6

To Find out  of sample  by using UV-VIS Spectrophotometer

7

 

Determination of unknown concentration of pharmaceutical substance by using UV-Visible spectroscopy

8

Determination of ℅ content of sample using UV-Visible spectrophotometer.

 

 

                      INFRARED SPECTROSCOPY

9

Identification of unknown compound using IR Spectroscopy

 

 

      NUCLEAR MAGNETIC RESONANCE SPECTROSCOPY

 

10

Identification of unknown compound using NMR Spectroscopy. Spectrum study and different functional groups interpretation.

 

 

                                      MASS SPECTROMETRY              

11

Identification of unknown compound using mass Spectrometry also study of fragmentation and relative abundance.

                          

12

                           pH  MEASUREMENT

 

 Introduction

COLORIMETRY:

                        It is a method of analysis in which colored solutions are analyzed by measuring intensity of transmitted light.

Colorimetry is the use of the human eye to determine the concentration of colored species.

Spectrophotometry is the use of instruments to make the same measurements.  It extends the range of possible measurements beyond those that can be determined by the eye alone.

 

TRANSMITTANCE:

                         When light is passed through a solution, a portion of light is absorbed while rest is transmitted. This property is called transmittance.

                     T= I/I0                     T= I/I0

I= Intensity of transmitted light

I0= Intensity of incident light                             

ABSORBANCE:

Absorbance is log of inverse of transmittance

                      A=log 1/T

The lamp emits all colors of light (i.e., white light). 

The monochromator selects one wavelength and that wavelength is sent through the sample. 

The detector detects the wavelength of light that has passed through the sample.

The amplifier increases the signal so that it is easier to read against the background noise.

 

 

Visual  Observations – Because colorimetry is based on inspection of materials with the human eye, it is necessary to review aspects of visible light.

Visible light is the narrow range of electromagnetic waves with the wavelength of 400-700 nm

 

 

Observed Color of Compound

 

Color of Light Absorbed

 

Approximate Wavelength of Light Absorbed

 

Green

 

Red

 

700 nm

 

Blue-green

 

Orange-red

 

600 nm

 

Violet

 

Yellow

 

550 nm

 

Red-violet

 

Yellow-green

 

530 nm

 

Red

 

Green

 

500 nm

 

Orange

 

Blue

 

450 nm

 

Yellow

 

Violet

 

400 nm

 

For more than one color:  the ratio of an unknown mixture can also be determined by matching the shade of the color to those produced from known ratios. The ratio of a mixture of red and blue can be determined visibly by comparing the mixture to purples produced from known ratios of red and blue.

 

 

To find out  of Potassium Per Magnate by using colorimeter.

 

EQUIPMENT:

Colorimeter

SAMPLE:

 Potassium per magnate

PROCEDURE:

·         Turn  on equipment few min prior to use

·         Place blank or reference solution in sample holder

·         Adjust absorbance to zero, using monochromator or filter of lowest wavelength.

·         Now replace blank solution with colored solution and note absorbance.

·         Repeat the same procedure for all the available filters (405,492,546,578,660nm).

·         Observe wavelength at which maximum absorbance obtained, which is considered to be

 

 

 

 

 

 

 

 

 

 

 

 

 

 

To find out  of Potassium Di Chromate by using colorimeter.

EQUIPMENT:

Colorimeter

SAMPLE:

 Potassium di chromate

PROCEDURE:

·         Turn  on equipment few min prior to use

·         Place blank or reference solution in sample holder

·         Adjust absorbance to zero, using monochromator or filter of lowest wavelength.

·         Now replace blank solution with colored solution and note absorbance.

·         Repeat the same procedure for all the available filters (405,492,546,578,660nm).

·         Observe wavelength at which maximum absorbance obtained, which is considered to be

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

  To find out unknown concentration of sample using colorimeter.

 

EQUIPMENT:

                Colorimeter

SAMPLE:

            Different concentrations of sample (e.g 1, 2, 3, 4 and 5 Potassium per magnate)

            Unknown concentration of sample

 

PROCEDURE:

·         Turn  on equipment few min prior to use

·         Place blank or reference solution in sample holder

·         Adjust absorbance to zero, using monochromator or filter of lowest wavelength.

·         Now replace blank solution with colored solution and note absorbance.

·         Since we are using KMnO4, we make different concentration and repeat above mentioned procedure on all of them and measure absorbance.

·         Now take solution of unknown concentration and check its absorbance.

·         Now by plotting values in graph we can determine concentration of unknown solution.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

                             UV-VISIBLE SPECTROSCOPY

 

Ultraviolet–visible spectroscopy or ultraviolet-visible spectrophotometry (UV-Vis or UV/Vis) refers to absorption spectroscopy or reflectance spectroscopy in the ultraviolet-visible spectral region. This means it uses light in the visible and adjacent (near-UV and near-infrared (NIR)) ranges. The absorption or reflectance in the visible range directly affects the perceived color of the chemicals involved. In this region of the electromagnetic spectrum, molecules undergo electronic transitions. This technique is complementary to fluorescence spectroscopy, in that fluorescence deals with transitions from the excited state to the ground state, while absorption measures transitions from the ground state to the excited state.

Principle of ultravoilet-visible absorption:

Molecules containing π-electrons or non-bonding electrons (n-electrons) can absorb the energy in the form of ultraviolet or visible light to excite these electrons to higher anti-bonding molecular orbitals. The more easily excited the electrons (i.e. lower energy gap between the HOMO and the LUMO), the longer the wavelength of light it can absorb.

    HOMO: High occupied molecular orbital

    LUMO: Lowest unoccupied molecular orbital

 

Beer-Lambert Law:

The method is most often used in a quantitative way to determine concentrations of an absorbing species in solution, using the Beer-Lambert law:

where A is the measured absorbance, in Absorbance Units (AU),  I0 is the intensity of the incident light at a given wavelength, I is the transmitted intensity, L the pathlength through the sample, and c the concentration of the absorbing species. For each species and wavelength, ε is a constant known as the molar absorptivity or extinction coefficient.

  

 

 

 

 

 

SCHEMATIC DIAGRAM DOUBLE BEAM UV-VIS SPECTROPHOTOMETER:

 

Instruments for measuring the absorption of U.V. or visible radiation are made up of the following components;

 ·  Sources (UV and visible)

 ·  Wavelength selector (monochromator)

 ·  Sample containers

·         Detector

·          Signal processor and readout

  

 

 

 

 

 

To Find out  of sample  by using UV-VIS Spectrophotometer

 

EQUIPMENT:

UV-VIS Spectrophotometer.

SAMPLE:

                 Potassium  dichromate

Procedure:

·         Turn  on equipment few min prior to use

·         Make 0.01 w/v solution of K2Cr2O7.

·         Fill a cuvette with water and other with sample solution

·         Set incident wavelength at 330nm

·         Bring water in liquid light path and blank it.

·         Get colored compound solution in light path and note the absorbance.

·         Repeat same procedure by increasing wavelength to 10 until wavelength reaches the value of 440nm.

·         Note the values of absorbance at various wavelengths and plot a graph of wavelength versus absorbance.

·         Determine  of sample.

 

 

 

 

 

 

 

 

 

 

Determination of unknown concentration of pharmaceutical substatance by using UV-Visible spectroscopy.

 

EQUIPMENT:

                      UV-Visible spectrophotometer.

PROCEDURE:                                                                                        

 PREPARATION OF STANDARD SOLUTIONS:

Prepare various standard solutions of colored compounds dissolving in 100ml of water. Label the beaker containing the solution.

OPERATION OF UV-VISIBLE  SPECTROPHOTOMETER:

·         Turn the spectrophotometer ON.

·         Place in cuvettes containing blank solution and colored compounds  of various concentrations.

·         Determine  of the compound for any concentration.

·         Now take readings of absorbance for all standard solutions.

·         Replace standard solution with sample solution of unknown concentration and note the absorbance.

 

PLOTTING A GRAPH AND ADMINISTRATION OF UNKNOWN CONCENTRATION:               

·         Make a graph of concentration (x-axis) versus absorbance (y-axis).

·         Plot the values of standard solutions and test solution.

·         Find the concentration value corresponding to the absorbance shown by the unknown solution via calibration curve method.

 

 

 

 

 

 

 

 

 

 

 

Determination of ℅ content of sample using UV-Visible spectrophotometer.

 

EQUIPMENT:

                     UV-Visible spectrophotometer      

SAMPLE:

                    Tetracycline capsule containing 250mg tetracycline hydrochloride

ASSAY:

         Mix the contents of twenty capsules, taking care not to lose any material when opening the capsule shells. Weigh the total contents of twenty capsules and calculate the average amount of powder containing 250mg of tetracycline hydrochloride. Take this accurately weighed quantity of powder equivalent to the average weight supposed to contain 0.250g antibiotic.Transfer this amount into 250ml volumetric flask, add about 100ml of 0.1N HCl, shake thoroughly and make up the volume with 0.1N HCl. Tranfer 10ml of this solution into a 100ml flask and make up the volume with water. Then transfer 10ml of the preceding solution into a 100ml flask, add 10ml of 0.1N NaOH and make up the volume with water. Measure the absorbance of a 1cm layer of the sample at  of 380. Calculate the concentration using a value of 380 as E (1 percent, 1cm).

 

Calculation through dilution factor:

Stated potency of capsule= 250mg

Dilution factor= 250/250 × 10/100 × 10/100 = 1/100

content of tetracycline hydrochloride in capsule

Absorbance of sample/ E (1, 1cm)  × 1000 × 100

Absorbance of sample/380 ×1000 ×100 = X

Therefore, content of tetracycline hydrochloride.

 

 

 

                             INFRARED SPECTROSCOPY

 

Infrared spectroscopy (IR spectroscopy) is the spectroscopy that deals with the infrared region of the electromagnetic spectrum that is light with a longer wavelength and lower frequency than visible light. It covers a range of techniques, mostly based on absorption spectroscopy. As with all spectroscopic techniques, it can be used to identify and study chemicals. A common laboratory instrument that uses this technique is a Fourier transform infrared (FTIR) spectrometer.

The infrared portion of the electromagnetic spectrum is usually divided into three regions; the near-, mid- and far- infrared, named for their relation to the visible spectrum. The higher-energy near-IR, approximately 14000–4000 cm−1 (0.8–2.5 μm wavelength) can excite overtone or harmonic vibrations. The mid-infrared, approximately 4000–400 cm−1 (2.5–25 μm) may be used to study the fundamental vibrations and associated rotational-vibrational structure. The far-infrared, approximately 400–10 cm−1 (25–1000 μm), lying adjacent to the microwave region, has low energy and may be used for rotational spectroscopy. The names and classifications of these subregions are conventions, and are only loosely based on the relative molecular or electromagnetic properties.

 

PRINCIPLE OF IR ABSORPTION:

 

IR radiation does not have enough energy to induce electronic transitions as seen with UV. Absorption of IR is restricted to compounds with small energy differences in the possible vibrational and rotational states.

For a molecule to absorb IR, the vibrations or rotations within a molecule must cause a net change in the dipole moment of the molecule. The alternating electrical field of the radiation (remember that electromagnetic radiation consists of an oscillating electrical field and an oscillating magnetic field, perpendicular to each other) interacts with fluctuations in the dipole moment of the molecule. If the frequency of the radiation matches the vibrational frequency of the molecule then radiation will be absorbed, causing a change in the amplitude of molecular vibration.

 

 

Molecular vibrations

The positions of atoms in a molecule are not fixed; they are subject to a number of different vibrations. Vibrations fall into the two main categories of stretching and bending.

Stretching: Change in inter-atomic distance along bond axis

 

 

Bending: Change in angle between two bonds. There are four types of bend:

  • Rocking
  • Scissoring
  • Wagging
  • Twisting

 

 

 

 

 

 

 

 

 

 

 

INSTRUMENTATION:

 

 

 

Identification of unknown compund using IR Spectroscopy

Typical Infrared Absorption Frequencies

Stretching Vibrations

Bending Vibrations

Functional Class

Range (cm-1)

Intensity

Assignment

Range (cm-1)

Intensity

Assignment

Alkanes

2850-3000

str

CH3, CH2 & CH
2 or 3 bands

1350-1470
1370-1390
720-725

med
med
wk

CH2 & CH3 deformation
CH3 deformation
CH2 rocking

Alkenes

3020-3100
1630-1680

1900-2000

med
var

str

=C-H & =CH2 (usually sharp)
C=C (symmetry reduces intensity)

C=C asymmetric stretch

880-995
780-850
675-730

str
med
med

=C-H & =CH2
(out-of-plane bending)
cis-RCH=CHR

Alkynes

3300
2100-2250

str
var

C-H (usually sharp)
C
C (symmetry reduces intensity)

600-700

str

C-H deformation

Arenes

3030
1600 & 1500

var
med-wk

C-H (may be several bands)
C=C (in ring) (2 bands)
(3 if conjugated)

690-900

str-med

C-H bending &
ring puckering

Alcohols & Phenols

3580-3650
3200-3550
970-1250

var
str
str

O-H (free), usually sharp
O-H (H-bonded), usually broad
C-O

1330-1430
650-770

med
var-wk

O-H bending (in-plane)
O-H bend (out-of-plane)

Amines

3400-3500 (dil. soln.)
3300-3400 (dil. soln.)
1000-1250

wk
wk
med

N-H (1°-amines), 2 bands
N-H (2°-amines)
C-N

1550-1650
660-900

med-str
var

NH2 scissoring (1°-amines)
NH2 & N-H wagging
(shifts on H-bonding)

Aldehydes & Ketones

2690-2840(2 bands)
1720-1740
1710-1720

1690

1675

1745

1780

med
str
str

str
str
str
str

C-H (aldehyde C-H)
C=O (saturated aldehyde)
C=O (saturated ketone)

aryl ketone
α, β-unsaturation
cyclopentanone
cyclobutanone


1350-1360
1400-1450
1100


str
str
med


α-CH3 bending
α-CH2 bending
 C-C-C bending

Carboxylic Acids & Derivatives

2500-3300 (acids) overlap C-H
1705-1720 (acids)
1210-1320 (acids)

1785-1815 ( acyl halides)

1750 & 1820 (anhydrides)

    1040-1100

1735-1750 (esters)

    1000-1300

1630-1695(amides)

str
str
med-str

str
str
str
str
str
str

O-H (very broad)
C=O (H-bonded)
O-C (sometimes 2-peaks)

C=O
C=O (2-bands)
    O-C
C=O
    O-C (2-bands)
C=O (amide I band)

1395-1440







1590-1650
1500-1560

med







med
med

C-O-H bending







N-H (1¡-amide) II band
N-H (2¡-amide) II band

Nitriles

Isocyanates,Isothiocyanates,
Diimides, Azides & Ketenes

2240-2260

2100-2270

med

med

CN (sharp)

-N=C=O, -N=C=S
-N=C=N-, -N
3, C=C=O

 

 

 

                          

 

 

                                  

 

 

                                  

 

 

 

 

 

                                 

 

     

 

      NUCLEAR MAGNETIC RESONANCE SPECTROSCOPY

 

Nuclear magnetic resonance spectroscopy, most commonly known as NMR spectroscopy, is a research technique that exploits the magnetic properties of certain atomic nuclei. It determines the physical and chemical properties of atoms or the molecules in which they are contained. It relies on the phenomenon of nuclear magnetic resonance and can provide detailed information about the structure, dynamics, reaction state, and chemical environment of molecules.

When placed in a magnetic field, NMR active nuclei (such as 1H or 13C) absorb electromagnetic radiation at a frequency characteristic of the isotope. The resonant frequency, energy of the absorption, and the intensity of the signal are proportional to the strength of the magnetic field.

Nuclear spin and the splitting of energy levels in a magnetic field

Subatomic particles (electrons, protons and neutrons) can be imagined as spinning on their axes. In many atoms (such as 12C) these spins are paired against each other, such that the nucleus of the atom has no overall spin. However, in some atoms (such as 1H and 13C) the nucleus does possess an overall spin. The rules for determining the net spin of a nucleus are as follows;

  1. If the number of neutrons and the number of protons are both even, then the nucleus has NO spin.
  2. If the number of neutrons plus the number of protons is odd, then the nucleus has a half-integer spin (i.e. 1/2, 3/2, 5/2)
  3. If the number of neutrons and the number of protons are both odd, then the nucleus has an integer spin (i.e. 1, 2, 3)

Chemical shift

The magnetic field at the nucleus is not equal to the applied magnetic field; electrons around the nucleus shield it from the applied field. The difference between the applied magnetic field and the field at the nucleus is termed the nuclear shielding.

Consider the s-electrons in a molecule. They have spherical symmetry and circulate in the applied field, producing a magnetic field which opposes the applied field. This means that the applied field strength must be increased for the nucleus to absorb at its transition frequency. This upfield shift is also termed diamagnetic shift.

Electrons in p-orbitals have no spherical symmetry. They produce comparatively large magnetic fields at the nucleus, which give a low field shift. This "deshielding" is termed paramagnetic shift.

 

Spin - spin coupling

Consider the structure of ethanol;

The 1H NMR spectrum of ethanol (below) shows the methyl peak has been split into three peaks (a triplet) and the methylene peak has been split into four peaks (a quartet). This occurs because there is a small interaction (coupling) between the two groups of protons. The spacing between the peaks of the methyl triplet are equal to the spacings between the peaks of the methylene quartet. This spacing is measured in Hertz and is called the coupling constant, J.

 

To see why the methyl peak is split into a triplet, let's look at the methylene protons. There are two of them, and each can have one of two possible orientations (aligned with or opposed against the applied field). This gives a total of four possible states;

In the first possible combination, spins are paired and opposed to the field. This has the effect of reducing the field experienced by the methyl protons; therefore a slightly higher field is needed to bring them to resonance, resulting in an upfield shift. Neither combination of spins opposed to each other has an effect on the methyl peak. The spins paired in the direction of the field produce a downfield shift. Hence, the methyl peak is split into three, with the ratio of areas 1:2:1.

Similarly, the effect of the methyl protons on the methylene protons is such that there are eight possible spin combinations for the three methyl protons;

Out of these eight groups, there are two groups of three magnetically equivalent combinations. The methylene peak is split into a quartet. The areas of the peaks in the quartet have the ration 1:3:3:1.

 

 

STRUCTURAL ELUCIDATION USING NMR SPECTROSCOPY

The note is that structure system is A3M2X2. Ha and Hx has the triplet pattern by Hm because of N+1 rule. The signal of Hm is split into six peaks by Hx and Ha.

 

          1) the structure of 2-methyl propanal is drawn 

2) Figure out which protons are chemically equivalent, i.e., two methyl (-CH3) groups are chemical equivalent.

3) Chemical shift of each protons is predicted by 1H chemical shift ranges (Fig.1): chemical shift of methyl groups (Ha) : 1-2 ppm (?Ha=1.1 ppm); chemical shift of -CH- groups (Hb) moves to downfield due to effect on aldehyde groups:2-3ppm ( ?Hb=2.4 ppm); chemical shift of aldehyde groups (Hc):9-10 ppm (?Hc=9.6 ppm)

4) Splitting pattern is determined by (N+1) rule: Ha is split into two peaks by Hb(#of proton=1). Hb has the septet pattern by Ha (#of proton=6). Hc has one peak. (Note that Hc has doublet pattern by Hb due to vicinal proton-proton coupling.)

 

 

 

 

 

 

 

 

 

                                      MASS SPECTROMETRY   

Mass spectrometry (MS) is an analytical technique that produces spectra (singular spectrum) of the masses of the molecules comprising a sample of material. The spectra are used to determine the elemental composition of a sample, the masses of particles and of molecules, and to elucidate the chemical structures of molecules, such as peptides and other chemical compounds. Mass spectrometry works by ionizing chemical compounds to generate charged molecules or molecule fragments and measuring their mass-to-charge ratios.

In a typical MS procedure, a sample, which may be solid, liquid, or gas, is ionized. The ions are separated according to their mass-to-charge ratio The ions are detected by a mechanism capable of detecting charged particles. Signal processing results are displayed as spectra of the relative abundance of ions as a function of the mass-to-charge ratio. The atoms or molecules can be identified by correlating known masses to the identified masses or through a characteristic fragmentation pattern.

INSTRUMENTATION:

A mass spectrometer consists of three components: an ion source, a mass analyzer, and a detector. The ionizer converts a portion of the sample into ions. There are a wide variety of ionization techniques, depending on the phase (solid, liquid, gas) of the sample and the efficiency of various ionization mechanisms for the unknown species. An extraction system removes ions from the sample, which are then trajected through the mass analyzer and onto the detector. The differences in masses of the fragments allows the mass analyzer to sort the ions by their mass-to-charge ratio. The detector measures the value of an indicator quantity and thus provides data for calculating the abundances of each ion present.

 

 

 

Alcohol

An alcohol's molecular ion is small or non-existent.  Cleavage of the C-C bond next to the oxygen usually occurs.  A loss of H2O may occur as in the spectra below.

3-Pentanol (C5H12O) with MW = 88.15

 

 

Aldehyde

Cleavage of bonds next to the carboxyl group results in the loss of hydrogen (molecular ion less 1) or the loss of CHO (molecular ion less 29).

3-Phenyl-2-propenal (C9H8O) with MW = 132.16

 

 

         

 

 

Alkane

Molecular ion peaks are present, possibly with low intensity.  The fragmentation pattern contains clusters of peaks 14 mass units apart (which represent loss of (CH2)nCH3).

Hexane (C6H14) with MW = 86.18

 

              

 

 

 

Measurement of Ph of different samples using Ph meter.

 

EQUIPMENT:

                      Ph meter

SAMPLE:

                  Hydrochloric acid

                  Nitric acid

                  Sodium hydroxide

               Water

PROCEDURE:

·         First of all make sure that pH meter is clean and make it in working condition

·         Take the sample whose pH is to be determined in a beaker

·         Dip Ph meter, measure for few seconds i.e for 3.5 sec

·         Required measurement will be obtained

·         Repeat same procedure for other compounds and measure their Ph values.

·         Record values in an table

 

 

 

 

 

 

 

 

 

 

 

 

                                                   

                             

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

S.NO

PRACTICALS

1.

SEPARATION OF MIXTURE OF INK USING PAPER CHROMATOGRAPHY

2.

 SEPARATION OF MIXTURE OF SUBSTANCE USING 2-D CHROMATOGRAPHY

 

3.

SEPARATION OF MIXTURE USING CIRCULAR PAPER CHROMATOGRAPHY

 

4.

COMPARISON OF RF VALUE OF ASPIRIN AND SALICYLIC ACID USING TLC.

 

5.

SEPARATION OF BETA CAROTENE AND CHLOROPHYLL FROM SPINACH LEAVES BY COLUMN CHROMATOGRAPHY

 

6.

DETERMINATION OF VISCOSITY OF DIFFERENT PHARMACEUTICAL FLUIDS.

 

7.

USE OF POLARIMETER AND MEASUREMENT OF ANGLE OF ROTATION OF AN OPTICALLY ACTIVE SUBSTANCE.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

                                             CHROMATOGRAPHY

                                                                                           

Chromatography is a physical method of separation that distributes components to separate between two phases, one stationary (stationary phase), the other (the mobile phase) moving in a definite direction.

Chromatography may be preparative or analytical. The purpose of preparative chromatography is to separate the components of a mixture for more advanced use (and is thus a form of purification). Analytical chromatography is done normally with smaller amounts of material and is for measuring the relative proportions of analytes in a mixture.

Chromatography terms

 

·         The analyte is the substance to be separated during chromatography.

·          Analytical chromatography is used to determine the existence and possibly also the concentration of analyte(s) in a sample.

·         A bonded phase is a stationary phase that is covalently bonded to the support particles or to the inside wall of the column tubing.

·         A chromatogram is the visual output of the chromatograph. In the case of an optimal separation, different peaks or patterns on the chromatogram correspond to different components of the separated mixture.

·          A chromatograph is equipment that enables a sophisticated separation, e.g. gas chromatographic or liquid chromatographic separation.

·         The eluate is the mobile phase leaving the column.

·         The eluent is the solvent that carries the analyte.

·          An eluotropic series is a list of solvents ranked according to their eluting power.

·         An immobilized phase is a stationary phase that is immobilized on the support particles, or on the inner wall of the column tubing.

·         The mobile phase is the phase that moves in a definite direction. It may be a liquid (LC and Capillary Electrochromatography (CEC)), a gas (GC), or a supercritical fluid (supercritical-fluid chromatography, SFC). The mobile phase consists of the sample being separated/analyzed and the solvent that moves the sample through the column. In the case of HPLC the mobile phase consists of a non-polar solvent(s) such as hexane in normal phase or polar solvents in reverse phase chromotagraphy and the sample being separated. The mobile phase moves through the chromatography column (the stationary phase) where the sample interacts with the stationary phase and is separated.

·         Preparative chromatography is used to purify sufficient quantities of a substance for further use, rather than analysis

·         The retention time is the characteristic time it takes for a particular analyte to pass through the system (from the column inlet to the detector) under set conditions.

·         The sample is the matter analyzed in chromatography. It may consist of a single component or it may be a mixture of components. When the sample is treated in the course of an analysis, the phase or the phases containing the analytes of interest is/are referred to as the sample whereas everything out of interest separated from the sample before or in the course of the analysis is referred to as waste.

·          The solute refers to the sample components in partition chromatography.

·         The solvent refers to any substance capable of solubilizing another substance, and especially the liquid mobile phase in liquid chromatography.

·         The stationary phase is the substance fixed in place for the chromatography procedure. Examples include the silica layer in thin layer chromatography

·          The detector refers to the instrument used for qualitative and quantitative detection of analytes after separation.

 

 Techniques by chromatographic bed shape

 

  Column chromatography

Column chromatography is a separation technique in which the stationary bed is within a tube. The particles of the solid stationary phase or the support coated with a liquid stationary phase may fill the whole inside volume of the tube (packed column) or be concentrated on or along the inside tube wall leaving an open, unrestricted path for the mobile phase in the middle part of the tube (open tubular column). Differences in rates of movement through the medium are calculated to different retention times of the sample.

Planar chromatography

Planar chromatography is a separation technique in which the stationary phase is present as or on a plane. The plane can be a paper, serving as such or impregnated by a substance as the stationary bed (paper chromatography) or a layer of solid particles spread on a support such as a glass plate (thin layer chromatography). Different compounds in the sample mixture travel different distances according to how strongly they interact with the stationary phase as compared to the mobile phase. The specific Retention factor (Rf) of each chemical can be used to aid in the identification of an unknown substance.

 

 

Paper chromatography

Paper chromatography is a technique that involves placing a small dot or line of sample solution onto a strip of chromatography paper. The paper is placed in a jar containing a shallow layer of solvent and sealed. As the solvent rises through the paper, it meets the sample mixture, which starts to travel up the paper with the solvent. This paper is made of cellulose, a polar substance, and the compounds within the mixture travel farther if they are non-polar. More polar substances bond with the cellulose paper more quickly, and therefore do not travel as far.

Thin layer chromatography

Thin layer chromatography (TLC) is a widely employed laboratory technique and is similar to paper chromatography. However, instead of using a stationary phase of paper, it involves a stationary phase of a thin layer of adsorbent like silica gel, alumina, or cellulose on a flat, inert substrate. Compared to paper, it has the advantage of faster runs, better separations, and the choice between different adsorbents. For even better resolution and to allow for quantification, high-performance TLC can be used.

 

Techniques by physical state of mobile phase

 

Gas chromatography

Gas chromatography is based on a partition equilibrium of analyte between a solid or viscous liquid stationary phase (often a liquid silicone-based material) and a mobile gas (most often helium). The stationary phase is adhered to the inside of a small-diameter (commonly 0.53 - 0.18mm inside diameter) glass or fused-silica tube (a capillary column) or a solid matrix inside a larger metal tube (a packed column). It is widely used in analytical chemistry; though the high temperatures used in GC make it unsuitable for high molecular weight biopolymers or proteins (heat denatures them), frequently encountered in biochemistry, it is well suited for use in the petrochemical, environmental monitoring and remediation, and industrial chemical fields. It is also used extensively in chemistry research.

Liquid chromatography

Liquid chromatography (LC) is a separation technique in which the mobile phase is a liquid. Liquid chromatography can be carried out either in a column or a plane. Present day liquid chromatography that generally utilizes very small packing particles and a relatively high pressure is referred to as high performance liquid chromatography (HPLC).

Affinity chromatography:

 

Affinity chromatography  is based on selective non-covalent interaction between an analyte and specific molecules. It is very specific, but not very robust. It is often used in biochemistry in the purification of proteins bound to tags. These fusion proteins are labeled with compounds such as His-tags, biotin or antigens, which bind to the stationary phase specifically. After purification, some of these tags are usually removed and the pure protein is obtained.

 

Special techniques:

 

Reversed-phase chromatography:

Reversed-phase chromatography (RPC) is any liquid chromatography procedure in which the mobile phase is significantly more polar than the stationary phase. It is so named because in normal-phase liquid chromatography, the mobile phase is significantly less polar than the stationary phase. Hydrophobic molecules in the mobile phase tend to adsorb to the relatively hydrophobic stationary phase. Hydrophilic molecules in the mobile phase will tend to elute first. Separating columns typically comprise a C8 or C18 carbon-chain bonded to a silica particle substrate.

Two-dimensional chromatography:

In some cases, the chemistry within a given column can be insufficient to separate some analytes. It is possible to direct a series of unresolved peaks onto a second column with different physico-chemical (Chemical classification) properties. Since the mechanism of retention on this new solid support is different from the first dimensional separation, it can be possible to separate compounds that are indistinguishable by one-dimensional chromatography. The sample is spotted at one corner of a square plate,developed, air-dried, then rotated by 90° and usually redeveloped in a second solvent system.

 

                                              Paper chromatography

 

Paper chromatography is an analytical method technique for separating and identifying mixtures that are or can be coloured, especially pigments. This can also be used in secondary or primary colours in ink experiments. This method has been largely replaced by thin layer chromatography, however it is still a powerful teaching tool.

Double-way paper chromatography, also called two-dimensional chromatography, involves using two solvents and rotating the paper 90° in between. This is useful for separating complex mixtures of similar compounds, for example, amino acids. if a filter paper is used, it should be of a high quality paper. The mobile phase is developing solutions that can travel up to the stationary phase carrying the sample alongside with it.

 

 

Types of Paper Chromatography

 

1.Descending Paper Chromatography-In this type development of chromatogram is done by allowing the solvent to travel down the paper is called Descending Chromatography. Here the mobile phase is present in the upper portion.

2.Ascending Paper Chromatography-Here the solvent travel upward direction of the Chromatographic paper. Both the Descending and Ascending Paper Chromatography are used for separation of Organic and Inorganic substances.

3.Ascending-Descending Paper Chromatography-It is the hybrid of both the above technique. The upper part of the Ascending chromatography can be folded over a rod and allowing the paper to become descending after crossing the rod.

4.Radial Paper Chromatography-It is also called as Circular chromatography. Here a circular filter paper is taken and the sample is given at the center of the paper. After drying the spot the filter paper tied horizontally on a Petridish containing solvent. So that Wick of the paper is dipped inside the solvent. The solvent rises through the wick and the component get separated in form of concentrate circular zone.

5.Two-Dimensional Paper Chromatography-In this technique a square or rectangular paper is used. Here the sample is applied to one of the corner and development is performed at right angle to the direction of first run.

 

 

RETENTION FACTOR:

The retention factor (Rƒ) may be defined as the ratio of the distance traveled by the substance to the distance traveled by the solvent. If Rƒ value of a solution is zero, the solute remains in the stationary phase and thus it is immobile. If Rƒ value = 1 then the solute has no affinity for the stationary phase and travels with the solvent front. To calculate the Rƒ value, take the distance traveled by the substance divided by the distance traveled by the solvent. For example, if a compound travels 2.1 cm and the solvent front travels 2.8 cm, (2.1/2.8) the Rƒ value = 0.75

 

SEPARATION OF MIXTURE OF INK USING PAPER CHROMATOGRAPHY

Apparatus:       

                Whatmann no.1 filter paper, chromatographic tank, capillary tube, pencil

Chemicals:

                Mixture of inks, red, blue, black and yellow ink

Procedure:

·         The spot of the test sample is loaded on the filter paper using a capillary tube. This spot should always be a concentrated but a very minute one.

·         Capillary tubes are used in paper chromatography, because a small quantity can be taken into the tube without any force.

·         The upper line (solvent front) is drawn on the paper from 2 cm on the top and the bottom line (base line) is drawn from 2 cm from the bottom of the paper.

·         Usually 0.01g of the sample is dissolved in the running solvent (1g). Micro liter quantities are used to spot on the paper by a capillary tube.

·         The diameter of the spot should be only up to few mili meters. The spotting should be done several times in order to get a concentrated spot.

·         This filter paper is then placed inside the chamber saturated with the solvent to develop the chromatogram.

·         The chromatogram is then heated in an oven to high temperatures. This can be done not only by an oven but also using a fan, air etc.

 

                                                                                              

 

SEPARATION OF MIXTURE OF SUBSTANCE  USING 2-D    CHROMATOGRAPHY

Double-way paper chromatography, also called two-dimensional chromatography, involves using two solvents and rotating the paper 90° in between. This is useful for separating complex mixtures of similar compounds, for example, amino acids. if a filter paper is used, it should be of a high quality paper. The mobile phase is developing solutions that can travel up to the stationary phase carrying the sample alongside with it.

Two way paper chromatography gets around the problem of separating out substances which have very similar Rf values.

PROCEDURE:

·         Cut out a 4.5̋̋̋̋̋̋ x 4.5cm filter paper.

·         Draw a line on filter paper using a pencil about 2 cm  from a baseline.

·         Apply small spots of mixture of substances and individual components to the pencil line.

·         Place the filter paper in the chromatographic tank.

·         Allow the chromatogram to develop until solvent front reaches around 1cm of the top.

·         Dry the chromatogram and turn out 90  ̊C in another beaker.

·         The component of mixture will separate into different components further owing to their presence for various components.

·         Calculate their Rf value.

 

 

 

 

 

 

SEPARATION OF MIXTURE USING CIRCULAR PAPER CHROMATOGRAPHY

Radial chromatography is a form of paper chromatography, a technique for separating colored substances from their mixtures.

Here the solvent travels from the center of the circular chromatography paper towards the periphery. The entire system is kept in a covered petri dish in order to develop a chromatogram.

The wick at the center of paper dips into mobile phase in a petridish by which the solvent drains on to the paper and moves the sample radially to form the sample spots of different compounds as concentric rings.

 

 


PROCEDURE:

·         Apply a small amount of mixture to be separated to the center of circular filter paper.

·         Dip the center of filter paper into petri dish and then cover it.

·         Allow the chromatogram to develop until the solvent front reaches about 1 inch short of paper border.

·         Remove the chromatogram from the petridish and allow it to air dry.

·         Calculate Rf value of each separated component.

 

 

                                                                                                                                                                     

                                                                  

 

 

 

 

                                                                           

 

 

                         THIN LAYER CHROMATOGRAPHY

Thin layer chromatography (TLC) is a chromatography technique used to separate non-volatile mixtures. Thin layer chromatography is performed on a sheet of glass, plastic, or aluminium foil, which is coated with a thin layer of adsorbent material, usually silica gel, aluminium oxide, or cellulose. This layer of adsorbent is known as the stationary phase.

After the sample has been applied on the plate, a solvent or solvent mixture (known as the mobile phase) is drawn up the plate via capillary action. Because different analytes ascend the TLC plate at different rates, separation is achieved.

Thin layer chromatography can be used to monitor the progress of a reaction, identify compounds present in a given mixture, and determine the purity of a substance.

 


 

PRINCIPLE:

Similar to other chromatographic methods TLC is also based on the principle of separation. The separation depends on the relative affinity of compounds towards stationary and mobile phase. The compounds under the influence of mobile phase (driven by capillary action) travel over the surface of stationary phase. During this movement the compounds with higher affinity to stationary phase travel slowly while the others travel faster. Thus separation of components in the mixture is achieved.

PLATE PREPARATION:

TLC plates are usually commercially available, with standard particle size ranges to improve reproducibility. They are prepared by mixing the adsorbent, such as silica gel, with a small amount of inert binder like calcium sulfate (gypsum) and water. This mixture is spread as thick slurry on an un reactive carrier sheet, usually glass, thick aluminum foil, or plastic. The resultant plate is dried and activated by heating in an oven for thirty minutes at 110 °C. The thickness of the absorbent layer is typically around 0.1 – 0.25 mm for analytical purposes and around 0.5 – 2.0 mm for preparative TLC.

 

COMPARISON OF RF VALUE OF ASPIRIN AND SALICYLIC ACID USING TLC.

CHEMICALS:

                 Aspirin and salicylic acid

MOBILE PHASE:

                 Acetone: ether (1:1)

 

 

 procedure:

A thin mark is made at the bottom of the plate with a pencil to apply the sample spots.

Then samples solutions are applied on the spots marked on the line at equal distances.

The mobile phase is poured into the TLC chamber to a level few centimeters above the chamber bottom

Then the plate prepared with sample spotting is placed in TLC chamber such that the side of the plate with sample line is towards the mobile phase. Then the chamber is closed with a lid.

The plate is immersed such that sample spots are well above the level of mobile phase but not immersed in the solvent as shown in the picture for development.

Allow sufficient time for development of spots. Then the plates are removed and allowed to dry. The sample spots are visualized in suitable UV light chamber .

 

QUESTION:

Which one will migrate more with non polar mobile phase? Why? Justify your answer.

 

Advantages of TLC

The Thin layer chromatography advantages include:

It is simple process with short development time.

It helps in visualization of separated compound spots easily.

The method helps to identify the individual compounds.

It helps in isolation of most of the compounds.

The separation process is faster and the selectivity for compounds is higher (even small differences in chemistry is enough for clear separation.

The purity standards of the given sample can be assessed easily.

It is a cheaper chromatographic technique

 

                                    COLUMN CHROMATOGRAPHY

 

Column chromatography is a method used to purify individual chemical compounds from mixtures of compounds. It is often used for preparative applications on scales from micrograms up to kilograms. The main advantage of column chromatography is the relatively low cost and disposability of the stationary phase used in the process. The latter prevents cross-contamination and stationary phase degradation due to recycling.

The classical preparative chromatography column, is a glass tube with a diameter from 5 mm to 50 mm and a height of 5 cm to 1 m with a tap and some kind of a filter (a glass frit or glass wool plug – to prevent the loss of the stationary phase) at the bottom. Two methods are generally used to prepare a column: the dry method, and the wet method.

·         For the dry method, the column is first filled with dry stationary phase powder, followed by the addition of mobile phase, which is flushed through the column until it is completely wet, and from this point is never allowed to run dry.

·         For the wet method, a slurry is prepared of the eluent with the stationary phase powder and then carefully poured into the column. Care must be taken to avoid air bubbles. A solution of the organic material is pipetted on top of the stationary phase. This layer is usually topped with a small layer of sand or with cotton or glass wool to protect the shape of the organic layer from the velocity of newly added eluent. Eluent is slowly passed through the column to advance the organic material. Often a spherical eluent reservoir or an eluent-filled and stoppered separating funnel is put on top of the column.

 

 

 

 

SEPARATION OF BETA CAROTENE AND CHLOROPHYLL FROM SPINACH  LEAVES  BY  COLUMN CHROMATOGRAPHY

Introduction

In this experiment, you will use column chromatography to separate and isolate two

colored compounds from spinach leaves. The first is beta carotene, which is the precursor to vitamin A, and has a yellow color. The second is chlorophyll, which plants use in the process of photosynthesis, and has a green color.

 

Column chromatography operates on the same principles as TLC, but instead of using a plate coated with silica gel, you will use a buret filled with alumina. Solvent will travel down the column instead of up a plate. The same concepts of polarity apply.

 

Beta carotene is a hydrocarbon, making it quite nonpolar. Chlorophyll contains very

polar bonds to magnesium as well as a few polar functional groups. This large difference in polarity makes this separation very effective. The beta carotene moves much more easily down the column than the chlorophyll.

 

Both beta carotene and chlorophyll are colored, which will make it easy to observe their movement down the column. The color comes from light being absorbed by the compounds because of their numerous C=C bonds.

 

 

Procedure

Separate the beta carotene and chlorophyll from the spinach leaves:

Obtain 10-14 spinach leaves. Pinch off the stems and wash off any dirt. Pat them as dry

as you can with paper towels. Then add leaves to your pile until you have at least 10 g of

leaves.

Place the leaves in a large mortar and add about 20 ml of dichloromethane and 20 ml of

petroleum ether. Mush the leaves carefully with a pestle for about 5 minutes to extract all

organic-soluble components.

Carefully pour off the organic solvents into a 100 or 50 ml round bottom flask, leaving any water (which will be at the bottom) behind.

Add about 500 mg of alumina to the flask. Rotovap to dryness, as described in

“Evaporating a Solution.”

 

Separate the beta carotene and chlorophyll from each other:

 

                   •     Prepare a column as described in "Separating Compounds by Column  

                           Chromatography."

·         Pour the sample plus alumina in the top of the column and wash it down with a few ml of hexanes if it sticks to the sides.

·         Prepare a solution of 10% ethyl acetate in hexanes (start with 100 ml and     

                            prepare more as needed).

                           

·         Fill the column with solvent, open the stopcock, and begin pushing it through with the bulb. At this point a yellow band should appear and move down the column – this is the beta carotene.

 

·         Collect it in a separate flask as it comes out the bottom (start

                                         collecting when the yellow band is about an inch from the

                                         bottom) when the yellow band has been  completely eluted

                                          push out any extra solvent.

 

                                 •     Prepare 100 ml of 10% methanol in dichloromethane.

 

·         Add the new solvent to the column and begin pushing it through a  green band should now move down the column.

 

                                  •     Collect this band in a separate flask as it is eluted from the column.

                                  •     Force all liquid from the column and leave it on my desk. (It is very           

                                         difficult to clean it out until it dries.)

                                                                                                             

          QUESTIONS:

 

·         Which is more polar, beta carotene or chlorophyll? If you didn’t have access to their structures, how would you know this from the experiment?

                                               

·         Which solvent system is more polar, 10% ethyl acetate in  

                                                    hexanes, or 10% methanol in dichloromethane? How can you

                                                    tell from the experiment?

                                                                                                  

 

 

                                                       VISCOSITY

viscosity is the quantity that describes a fluid's resistance to flow. Fluids resist the relative motion of immersed objects through them as well as to the motion of layers with differing velocities within them.

Viscosity is the measure of the internal friction of a fluid. This friction becomes apparent when a layer of fluid is made to move in relation to another layer. The greater the friction, the greater the amount of force required to cause this movement, which is called shear. Shearing occurs whenever the fluid is physically moved or distributed, as in pouring, spreading, spraying, mixing, etc. Highly viscous fluids, therefore, require more force to move than less viscous materials.

 

DETERMINATION OF VISCOSITY OF DIFFERENT PHARMACEUTICAL FLUIDS.

BROOK FIELD VISCOMETER:

The rotational viscometer is based upon the measurement of torque of a rotating spindle a sample at a specified velocity.

The instrument based on the fact that solid rotating body immersed in a liquid is subjected to the retarding force due to the viscous drag.

The force is directly proportional to the viscosity of the substance.

Apparatus:

Brook field viscometer

Chemicals:

Glycerol, Chloroform, acetone, ether, 70℅ sucrose.

Procedure:

Adjust the bubble level at the top of the instrument.

Introduce the medicinal compound in a beaker.

Place spindle of suitable size.

Adjust the spindle number and the speed required.

Turn on the viscometer and note the viscosity.

Repeat the procedure on all the fluids.

The obtained values can be compared with literature values to determine the viscosity of a substance.

 

 

 

 

 

 USE OF POLARIMETER AND MEASUREMENT OF ANGLE OF ROTATION OF AN OPTICALLY ACTIVE SUBSTANCE.

 

A polarimeter is a scientific instrument used to measure the angle of rotation caused by passing polarized light through an optically active substance.

Some chemical substances are optically active, and polarized (unidirectional) light will rotate either to the left (counter-clockwise) or right (clockwise) when passed through these substances. The amount by which the light is rotated is known as the angle of rotation.

 

Construction

The polarimeter is made up of two Nicol prisms (the polarizer and analyzer). The polarizer is fixed and the analyzer can be rotated. The prisms may be compared to as slits S1 and S2. The light waves may be considered to correspond to waves in the string. The polarizer S1 allows only those light waves which move in a single plane. This causes the light to become plane polarized. When the analyzer is also placed in a similar position it allows the light waves coming from the polarizer to pass through it. When it is rotated through the right angle no waves can pass through the right angle and the field appears to be dark. If now a glass tube containing an optically active solution is placed between the polarizer and analyzer the light now rotates through the plane of polarization through a certain angle, the analyzer will have to be rotated in same angle.

Operation

Polarimeters measure this by passing monochromatic light through the first of two polarising plates, creating a polarized beam. This first plate is known as the polarizer. This beam is then rotated as it passes through the sample. The sample is usually prepared as a tube where the optically active substance is dissolved in an optically inactive chemical such as distilled water, ethanol, methanol. Some polarimeters can be fitted with tubes that allow for sample to flow through continuously.

After passing through the sample, a second polarizer, known as the analyzer, rotates either via manual rotation or automatic detection of the angle. When the analyzer is rotated to the proper angle, the maximum amount of light will pass through and shine onto a detector.

 

 

 

 

 

 

 

 

Apparatus:

              Polari meter,  glass tube.

Chemicals:

              D-Glucose.

 

Procedure:

Turn on polarimeter and allow to heat for 5min

Make a 1℅ w/v D-Glucose solution and fill in the 100mm test tube provided along with  the polarimeter.

Check for zero position of dial. Incase of an error, add or subtract error value during procedure.

Ideally at zero position viewing field should show a darker central region against a bright region.

Once viewing field is adjusted, open lens cover and put the test tube containing glucose solution in it.

Close the cover with bulb part of tube upward in order to store air bubbles in it.

Throughout the observation process adjust the dial such that viewing field become identical.

Read out angle of light rotation. Calculate specific rotation of unknown compound.

If all other parameters are known. The equipment can also be utilized to determine concentration of unknown substance specific rotation can be found by formula:

                                      

 In this equation, l is the path length in decimeters and c is the concentration in g/mL, for a sample at a temperature T (given in degrees Celsius) and wavelength λ (in nanometers). If the wavelength of the light used is 589 nanometer (the sodium D line), the symbol “D” is used. The sign of the rotation (+ or −) is always given. When using this equation, the concentration and the solvent may be provided in parentheses after the rotation. The rotation is reported using degrees, and no units of concentration are given (it is assumed to be g/100mL).

    

 

 

 

 


No comments:

Post a Comment